Ne215 counter connect – Baumer NE215 User Manual

Page 7

Advertising
background image

7

NE215

Counter connect

3.1 Connecting the power supply

For power supply following voltage ranges are available:

AC 115/230 VAC (50/60 Hz)

DC 24 VDC ±10%

Voltage supply

Recommended external fuse

115/230 VAC

M 125 mA

24 VDC

M 400 mA

Assign voltage supply to terminals 2 and 3 according to

diagramm.

DC-voltage 24 VDC:

Connect voltage supply that is free from any interference. Do not uti-

lize the voltage supply for parallel supply of drives, shields, magnetic

valves, etc. Supplying lines must be separated from lines providing

load current.

Fire protection: Operate the instrument using the recommended

external fusing indicated in the terminal diagram. EN 61010 specifies

that 8 A/150 VA (W) must never be exceeded in the event of a fault.

3.2 Assignment signal output „relay contact“

Terminals 4, 5, 6 and 7, 8, 9 are no-potential changeover contacts.

Terminals 10 and 11 are configured as NC or NO contacts in ac-

cordance with the purchase order specification. The signal outputs

can be assigned in accordance with the adjoining wiring diagram.

Implementation as a pulse or continuous signal, together with the

pulse time, is effected in the programming mode (lines 31, 32, 33).

Max. rating

Max. voltage

Max. current

150 VA/30 W

265 V

1 A

The user is responsible for ensuring that a switching load of 8 A/150

VA (W) is not exceeded in the event of a fault. Internal spark sup-

pression by means of two zinc oxide varistors (275 V).

Assign terminals 4, 5, 6; 7, 8, 9; and 10, 11 (relay contactoutputs)

accordingly.

6 5 4

7 8 9

11 10

115 VAC 230 VAC