3-4 Message Communication

3-23

3

C

om
m

uni

ca

tin

g

w

ith FD

-M

H

Se
ries

- EtherNet/IP Compatible Network Unit DL-EP1 User’s Manual (FD-MH) -

(2) Writing data to a sensor amplifier

Message communication writes parameters to the DL-EP1 and the sensor
amplifier connected to the DL-EP1.

"(2) Writing parameters to a sensor amplifier" (Page 3-44)

(3) Issuing operation commands to a sensor amplifier

Message communication issues operation commands such as integration reset to
the DL-EP1 and the sensor amplifier connected to the DL-EP1.

"(3) Issuing operation commands to a sensor amplifier" (Page 3-45)

(4) Locking sensor amplifier operations

Message communication locks operations by keys on the sensor amplifier
connected to the DL-EP1.

"(4) Locking sensor amplifier operations" (Page 3-47)

FD-MH

Command:
1000

Response

FD-MH
Integrated flow rate setting value 1: 1501000

DL-EP1

Scanner

<Details of command>
Service code: 10H
Class ID: 67H
Instance ID: 01H
Attribute ID: 0345H(837)
Service data: 1000

FD-MH

Command

Response

Integration reset execution by FD-MH

DL-EP1

Scanner

<Details of command>
Service code: 4BH
Class ID: 67H
Instance ID: 01H
Attribute ID: 0302H(770)

FD-MH

Command

Response

Changes using the
sensor amplifier keys are locked.

DL-EP1

Scanner

<Details of command>
Service code: 4CH
Class ID: 67H
Instance ID: 00H
Service data: 01H