Parameters of Commands and Responses

024

Calculated value three-point calibration
SET2 Confirmation operation request

*5*12*13

*

1

R

0-1: Last written value

1

W

0

o1: Perform calculated value three-

point calibration 2nd point
request

*3

025

Calculated three-point calibration
SET3 Confirmation operation request
(Perform calculated three-point
calibration.)

*5*12*13

*

1

R

0-1: Last written value

1

W

0

o1: Perform calculated three-point

calibration 3rd point request

*3

026

One-point tuning request for diff. count filter

*

1

R

0-1: Last written value

1

W

One-point tuning request for diff. count

filter

*3

027

Two-point tuning
1st point confirmation operation request for
diff. count filter

*

1

R

0-1: Last written value

1

W

Two-point tuning for diff. count filter

*3

028

Two-point tuning
2nd point confirmation operation request for
diff. count filter
(Determine HIGH and LOW setting value.)

*

1

R

0-1: Last written value

1

W

Two-point tuning for diff. count filter

*3

065

HIGH setting value (BANK 0)

±**.***

(±***.**)
((±****.*))

7

R/W

-99.999 to +99.999
(-999.99 to +999.99)
((-9999.9 to +9999.9))

5.000 (+50.00)

((+500.0))

066

LOW setting value (BANK 0)

7

R/W

-5.000 (-50.00)

((-500.0))

067

Shift target value (BANK 0)

7

R/W

0.000 (0.00)

((0.0))

068

Analog output - upper limit value
(BANK 0)

*5 *6

7

R/W

10.000 (100.00)

((1000.0))

069

Analog output - lower limit value
(BANK 0)

*5 *6

7

R/W

-10.000 (-100.00)

((-1000.0))

070

HIGH setting value (BANK 1)

7

R/W

5.000 (50.00)

((500.0))

071

LOW setting value (BANK 1)

7

R/W

-5.000 (-50.00)

((-500.0))

072

Shift target value (BANK 1)

7

R/W

0.000 (0.00)

((0.0))

073

Analog output - upper limit value
(BANK 1)

*5 *6

7

R/W

10.000 (100.00)

((1000.0))

074

Analog output - lower limit value
(BANK 1)

*5 *6

7

R/W

-10.000 (-100.00)

((-1000.0))

075

HIGH setting value (BANK 2)

7

R/W

5.000 (50.00)

((500.0))

076

LOW setting value (BANK 2)

7

R/W

-5.000 (-50.00)

((-500.0))

077

Shift target value (BANK 2)

7

R/W

0.000 (0.00)

((0.0))

078

Analog output - upper limit value
(BANK 2)

*5 *6

7

R/W

10.000 (100.00)

((1000.0))

079

Analog output - lower limit value
(BANK 2)

*5 *6

7

R/W

-10.000 (-100.00)

((-1000.0))

080

HIGH setting value (BANK 3)

7

R/W

5.000 (50.00)

((500.0))

081

LOW setting value (BANK 3)

7

R/W

-5.000 (-50.00)

((-500.0))

082

Shift target value (BANK 3)

7

R/W

0.000 (0.00)

((0.0))

083

Analog output - upper limit value
(BANK 3)

*5 *6

7

R/W

10.000 (100.00)

((1000.0))

084

Analog output - lower limit value
(BANK 3)

*5 *6

7

R/W

-10.000 (-100.00)

((-1000.0))

097

Key lock function

*

1

R/W

0: Not key locked state
1: Key locked state

0

098

Bank function

*7

*

1

R/W

0 to 3

0

099

Timing input

*8

*

1

R/W

0: Timing input OFF
1: Timing input ON

0

100

Laser emission stop input

*8

*

1

R/W

0: Emission stop input OFF
1: Emission stop input ON

0

Data

number

Data name

Data

type

*1

Number of
characters

Attribute

*2

Readout range

*14

Initial

value

*14

Parameters of Commands and Responses

104

Sub display's screen

*

1

R/W

0: R.V. value screen
1: Analog value screen
2: HI setting value screen
3: LO setting value screen
4: Zero shift value screen
5: CALC value screen

0

105

System parameter settings

*9

**

2

R/W

0 to 15

0

106

Tolerance tuning - tolerance setting range

±**.***

(±***.**)
((±****.*))

7

R/W

-99.999 to +99.999
(-999.99 to +999.99)
((-9999.9 to +9999.9))

5.000 (50.00)

((500.0))

107

Calibration function

*

1

R/W

0: Initial state
1: User setting

0

108

Calibration function SET1

±**.***

(±***.**)
((±****.*))

7

R/W

-99.999 to +99.999
(-999.99 to +999.99)
((-9999.9 to +9999.9))

0.000 (0.00)

((0.0))

109

Calibration function SET2

7

R/W

5.000 (50.00)

((500.0))

110

Calculated value calibration function

*5

*

1

R/W

0: Initial settings
1: Calculated two-point calibration
2: Calculated three-point calibration

0

111

Calculated value two-point calibration
function SET1

*5

±**.***

(±***.**)
((±****.*))

7

R/W

-99.999 to +99.999
(-999.99 to +999.99)
((-9999.9 to +9999.9))

5.000 (50.00)

((500.0))

112

Calculated value two-point calibration
function SET2

*5

7

R/W

10.000 (100.00)

((1000.0))

113

Calculated value three-point calibration
function SET1

*5

7

R/W

5.000 (50.00)

((500.0))

114

Calculated value three-point calibration
function SET3

*5

7

R/W

10.000 (100.00)

((1000.0))

129

Calculation function

*5

*

1

R/W

0: OFF
1: Addition mode
2: Subtraction mode

0

131

Measurement direction

*

1

R/W

0: nor
1: rEv

0

132

Sampling cycle

*

1

R/W

0: Initial settings
1: 0.33 ms
2: 1 ms
3: 2 ms
4: 5 ms

0

133

Averaging/Diff. count filter/High-pass filter

**

2

R/W

0: 1 time
1: 2 times
2: 4 times
3: 8 times
4: 16 times
5: 32 times
6: 64 times
7: 128 times
8: 256 times
9: 512 times
10: 1024 times
11: 2048 times
12: 4096 times
13: Diff. count filter
14: High-pass filter

4

134

Output mode

*

1

R/W

0: N.O. (Normally open)
1: N.C. (Normally closed)

0

136

Hold function setting

*

1

R/W

0: Sample hold
1: Peak hold
2: Bottom hold
3: Peak-to-peak hold
4: Auto peak hold
5: Auto bottom hold

0

137

Auto peak hold or auto bottom hold trigger
level

±**.***

(±***.**)
((±****.*))

7

R/W

-99.999 to +99.999
(-999.99 to +999.99)
((-9999.9 to +9999.9))

1.000 (10.00)

((100.0))

138

Timing input setting

*

1

R/W

0: Level
1: Edge

0

139

Delay timer

*

1

R/W

0: OFF
1: Delay ON
2: Delay OFF
3: One-shot

0

140

Timer duration

****

4

R/W

5 to 9999

60

Data

number

Data name

Data

type

*1

Number of
characters

Attribute

*2

Readout range

*14

Initial

value

*14