A: Calibrating with 2 master workpieces that have different dimensions

B: Calibrating with 1 master workpiece

3. Scaling settings

4. Click the button for “Send settings to controller”. (pg. 5 of this guide)

This will explain the individual setting methods for the following two patterns.
A: Calibrating with 2 master targets that have different dimensions
B: Calibrating with 1 master workpiece

* When measuring workpieces that have different thicknesses, accurate calibration can be performed with method A.

Once the “Measurement value acquisition start” button has been clicked, Measurement Value Acquisition/Display will appear.

Ex.) Setting scaling for OUT3 with 1 mm and 1.5 mm gauge blocks

(1) Click “Set from measurement data” in the scaling/offset settings for

OUT3.

(2) Insert the 1 mm gauge, click “Set” for input value 1 and then enter

1 mm for display value 1.

(3) Insert the 1.5 mm gauge, click “Set” for input value 2 and then enter

1.5 mm for display value 2.

(4) Click [OK] and then click “Send settings to controller”.

Ex.) Setting scaling for OUT3 with a 1 mm gauge

(1) Enter 1 mm for the scaling setting offset for OUT3.
(2) Click “Send settings to controller”.
(3) Measure the 1 mm master workpiece.
(4) Click “View Measurement value” and then click “Measurement value acquisition start”.
(5) Once “Zero” is clicked for OUT3, the thickness of the master workpiece will be displayed as the measurement value.
With the current settings “1.000” will be displayed.