KEYENCE LK-G5000 Series User Manual

Page 9

Advertising
background image

9

Measurement/Tolerance-Measurement type

Measurement/Tolerance-Filter

Measurement/Tolerance - Tolerance Evaluation

Selects which of the following targets will be measured in “Measurement type”.
Displacement/Speed/Acceleration

Settings related to measurement value processing are performed in Filter. By setting the moving average, it is possible to stabilise
measurement values.

Can set the evaluation values for the allowable measurement value range. The upper/lower limit values are set so that a signal is
output from the controller when the measurement values exceed these values.

The measurement values are set with a moving average in the range of 1 to 262144 times.
When an average of 4x has been set, the following process will occur.

Internal data

Outputs the average of Data 1 to 4 as Data A.

Outputs the average of Data 2 to 5 as Data B.

Outputs the average of Data 3 to 6 as Data C.

Sampling cycle

Average

measuring time

Average

measuring time

Average

measuring time