Basic Function Codes - The Modbus packet structure is determined by the function code being

performed. Some of the basic commands that CMS supports are shown in Table 4.

Table 4: Modbus function codes

Function

Hex Code

Read Holding Registers

03

Read Input Registers

04

Read Exception Status

07

Write Multiple Registers

10

Write Holding Register

06

Exception Packet Definition - If an error is detected in one of the following packet definitions a

certain error code is applied and sent back in an exception response.

Table 5: Modbus error codes

Code

Description

01

Invalid function code – function code not supported by device

02

Invalid data address – address defined by the start address and

number of registers is out of valid range

03

Invalid data value – number of registers = 0 or > 125

Table 6: Modbus exception response

Type

Byte Length

Example

Device address

1 Byte (1 – 247)

0x01

Function code

1 Byte (Function code + 0x80)

0x83

Exception code

1 Byte

0x01