Pressure/temperature (pt) probe, Thermocouple (tc) temperature sensors and arrays – Intek RheoVac CMS User Manual

Page 8

Advertising
background image

September 2013

4

© Intek, Inc. 2013

Revision D

1.3.3 Rheotherm Cooling Water Flow and Fouling (CWFF) Meters

Primary Calibration Accuracy:

Better than ±2% of reading

Flow Range:

2-20 ft/s (extended range optional)

Repeatability:

±0.5% of reading

Operating Temperature:

Electronics: 40 to 120°F (5 to 49°C)

Sensor: 40 to 140°F (5 to 60°C)

Operating Pressure:

60 psi (1000 psi optional)

Storage Temperature:

-20 to 140°F (-29 to 60°C)

Storage Pressure:

60 psig (maximum)

Wetted Surface:

300 Series SS (material options available)

Neoprene, Polyolefin (cable included)

1.3.4 Pressure/Temperature (PT) Probe

Accuracy:

±0.02 inches HgA for pressure

±0.1°F for temperature

Repeatability:

±0.5% of reading

Operating Temperature:

Electronics: 40 to 120°F (5 to 49°C)

Probe: 40 to 300°F (5 to 149°C)

Operating Pressure:

0.5 to 10 inches Hg absolute

15 psia maximum

Storage Temperature:

-20 to 185°F (-29 to 85°C)

Storage Pressure:

15 psig (maximum)

Process Connection:

Ball valve assembly (1½" thread-o-let must be welded to pipe for installation)

Wetted Surface:

300 Series SS

1.3.5 Thermocouple (TC) Temperature Sensors and Arrays

Temperature range:

40-175°F (higher temperature components optional)

Accuracy:

±0.2°F

Repeatability:

±0.1°F

Wetted Surface:

300 Series SS and thermoplastics

Advertising