4.2.4 Modbus Troubleshooting

Problem

Cause

Solution

No response from CMS

1. System not set for Modbus

communication

2. Incorrect Modbus settings
3. Improper wiring

1. Verify settings in serial.ini file, see Table 8
2. Verify wiring connections

Not enough data

resolution in Multiplied

Integer mode

1. Non-optimal multiplier values set

1. Change the multiplier values in the

MDBSInts.dat file

Multiplier values are

changing

1. The CMS has overflow protection

built in. When [process
variable]*[multiplier value]
exceeds 65535, the instrument
automatically lowers the multiplier
value to retain a usable output.

1. Request the multiplier value along with process

data rather than hard coding the multiplier value
in the client software

2. Lower the multiplier value for each process

variable that reaches the overflow state

3. Use single precision float mode instead of

multiplied integer mode