MESSAGE CODE

Description/Symptom

PROBABLE CAUSE

ACTION

OUTPUT ERROR

CODES

PT Probe Message Codes

MSG-2

Reserved

MSG-3

Reserved

MSG-4

Reserved

MSG-5 or (CalX)

Calibration Expired

1. Probe Calibration Expired

1. Contact factory for

Recalibration

N/A

Flow Meter Message Codes

MSG-2

Reserved

MSG-3

Reserved

MSG-4

Reserved

MSG-5

Sensor Issue

1. No sensor communication
2. Irregular sensor data

1. Contact factory

-101

Thermocouple Message Codes

Open

TC Circuit Open

1.

Loose connection

2.

Broken wire

1. Check wiring at the electronics
2. Contact factory

-500

Under

TC Value Under Range

1.

Temperature below 32°F

1. Contact factory

-501

Over

TC Value Over Range

1.

Temperature above 212°F

1. Contact factory

-502

Error

General Error

1.

Loose connection

2.

Component failure

1. Check wiring at the electronics
2. Contact factory

-504

MA (4-20mA DAQ) Message Codes

Open

MA Circuit Open

1. Loose connection
2. Broken wire

1. Check wiring at the electronics
2. Contact factory

-500

Under

MA Value Under Range

1. Output below range

1. Contact factory

-501

Over

MA Value Over Range

1. Output above range

2. Contact factory

-502

Error

General Error

1. Loose connection
2. Component failure

1. Check wiring at the electronics
2. Contact factory

-504