Calibration
• 75
6
Calibration
In data acquisition processes, how to calibrate your measurement devices to 
maintain its accuracy is very important. Users can calibrate the analog input 
and output channels under the users' operating environment to maximize the 
accuracy. This chapter will guide you though how to calibrate the PCI-9112. 
6.1 What do you need
Before calibrating your PCI-9112 card, you need to prepare the following 
equipment’s and materials for the calibration process: 
• Calibration program: Once the program is executed, it will guide you
through the calibration process. This program is included in the 
delivered package. 
• A 5 1/2 digit multimeter (6 1/2 is recommended)
• An adjustable voltage calibrator or a very stable and noise free DC
voltage generator. The calibrator should be able to provide voltage 
accuracy of up to 1/2 LSB. In bipolar 5V input range mode (GAIN = 
1), the voltage of 1/2 LSB is 1.22mV (i.e. (10V / 4096) / 2). When in 
unipolar 1.25V input range (GAIN = 8), the voltage of 1/2 LSB is 
0.153mV (i.e. (1.25V / 4096) / 2).