The power panel, Manikin setup – Laerdal SimMan 3G Mystic User Manual
Page 13
![background image](/manuals/738880/13/background.png)
9
10
Manikin Setup
System Setup
Setup f
or Sim
ulation
Clinical F
eatur
es
Maintenance
Tr
oubleshooting
Spar
e Par
ts
After connecting the batteries, connect the manikin to the external
power supply (12V to 24V) while the manikin is OFF.
The batteries will charge if the manikin is ON and connected to and
external power supply in the range of (20V - 24V)
Battery - Cable and Tube Descriptions
Name/Label
Tube/Cable Color
Connector Description
Battery 1
Black harness cable
Black rectangular
connector, 6 lead
Battery 2
Black harness cable
Black rectangular
connector, 6 lead
For more information on charging the batteries see section, Manikin
Setup, Battery Charging and Use.
The Power Panel
The power panel is found on the right side of the manikin, under a
loose skin flap. Lift the skin flap and pull out the protective cover.
To ensure easy access, use the zippered clothing provided with the
SimMan 3G manikin.
DC Input 9-24V X, XA
1
2
3
4
5
6
Power panel overview
1 Power ON / OFF button
2 Power status indicator
3 Battery status indicator
4 Charging status indicator
5 LAN network cable connector
6 External power supply connector
Power Panel LEDs Description
LED
Color
Power Status
Battery Status
Charge Status
Red
Power save*
0% - 20%
Not charging**
Yellow
Start up
20% - 70%
Charging
Green
Running
70% - 100%
Charge almost
complete***
No light
Off
Off
No charge****
* Blinking light
** One or both batteries missing, overheated,
damaged or otherwise not able to charge
*** Not recommended to charge the batteries too long
**** No power input, batteries are charged.
Power Save is activated when ever manikin is paused.
Manikin Setup