Maintenance – Laerdal SimMan 3G Mystic User Manual
Page 52
49
50
Maintenance
7 The Manikin Update program will start identifying software
versions currently installed. This will take about 2 minutes.
8 All programs and software versions are listed. The update status
for each program is displayed under <Update>.
9 Click <OK> to begin updating the software.
10 Complete the update wizard.
Note that some components might fail updating on first
attempt. Please ignore this and allow the update process
to continue.
The <SimMan3G> <Update Manager – Report> will show all
updates and show which ones were performed successfully or failed.
• Repeat from step 5 until all components are successfully
updated.
• Allow manikin time to save changes and reboot. Do not power
off manikin until it starts breathing again.
TROUBLESHOOTING
1 Should the following message appear on the screen during the
update, click <Yes> to continue the installing updates.
2 If a module fails to update, run the <Manikin Update> again.
Should the same module fail repeatedly, check to see whether
all internal wiring is connected and functioning properly.
ADVANCED VIDEO SERVER (AVS):
If you use an AVS with your SimMan system, please make sure that all
components are updated with the latest software.