Setup for Simulation

Manikin Setup

Cleaning the Fluid and Blood
System

After using the bleeding and or fluid system, the system must be
cleaned to prevent residue from clogging the valves and tubes. Keep
any wounds used connected while flushing the system with a cleaning
solution.

1 Ensure the manikin power is on.

2 Ensure the manikin’s internal reservoirs are drained. See Preparing

the Internal Reservoirs for filling.

3 Fill a bottle with a 60% Isopropanol solution or 70% Ethanol

solution.

4 Connect the bottle containing the alcohol solution to either the

blood or fluid fill connector and air connector to fill the manikin
with cleaning solution. Perform this action for both reservoirs. If
the manikin will be stored for long periods of time or transported
in freezing conditions, it is recommended to fill the reservoirs with
one of thee recommended alcohol solution to prevent damage
and preserve the integrity of the blood and fluid systems.

CLEANING THE FLUID SYSTEM

Note: To speed up the flushing process, follow the instruction for

Unclogging Fluid Tubes.

1 From the Instructors application, open <Circulation and fluids>

tab, check the Secretions box for urine, and then select polyuria.
Check the Secretions boxes for Sweat, Ears, Eyes, Nose and
Mouth.

2 Allow the Fluid system to flush with the cleaning solution until the

internal reservoir is empty.

3 When finished, uncheck all of the Secretions boxes.

CLEANING THE BLOOD SYSTEM

1 From the <Instructor Application>, open Circulation and fluids

tab, check the boxes for Upper port and Lower port, then select
Venous from the adjacent drop-down menus. Move the sliders
(to the right) to maximum bleeding rates.

2 Allow the Bleeding system to flush with the cleaning solution until

the internal reservoir is empty.

3 The blood system should be flushed repeatedly until the

discharged fluids are clear.

4 When finished, uncheck all of the boxes (in the Instructors

application).