Using the internal compressor, Turning the internal compressor off, Changing default compressor settings – Laerdal SimMan 3G Mystic User Manual

Page 28: Setup for simulation

Advertising
background image

25

26

Setup for Simulation

Using the Internal Compressor

The SimMan3G manikin’s chest movements, airway modes and
fluid systems are driven by compressed air. The right leg contains a
compressor and tank with separate reservoirs for clear and simulated
blood fluids.

For extended periods or stationary use, it is recommended to connect
to an external source of compressed air. This reduces wear on the
internal compressor and extends battery life of the manikin.

For instructions on connecting an external compressor and adjusting
compressor default settings see – Connecting external Air and CO

2

Supply.

To avoid overheating and reduce wear

− If using the manikin in temperatures above 40°C (104°F), always

allow the manikin to cool down between training sessions.

− If the manikin is placed in a bed, it should never be covered with

a comforter, duvet or quilt as these prevent heat transfer from
the manikin.

Turning the Internal Compressor
OFF

Switch OFF the internal compressor to conserve the manikin batteries
and reduce wear.

Turn OFF the internal compressor from the Instructor Application:
1 Select the <Tools> menu.

2 From <Manikin Setup>, click the Turn off internal compressor.

Changing Default Compressor
Settings

Change default compressor settings via the Profile Editor.

1 Open the Profile Editor from the <Tools> menu in Instructor

Application.

2 Select the <General> tab in the Profile Editor.

3 From the compressor setup option, choose the Internal or

External compressor as default.

Advertising