Troubleshooting – Laerdal SimMan 3G Mystic User Manual

Page 58

Advertising
background image

55

56

Troubleshooting

of the Instructor Application will indicate which manikin is pulse
palpated.

AIRWAY CONTAMINATION

Problem

• Manikin airways have become contaminated from mouth-to-

mouth rescue breathing

Solution

• Clean the outside of the manikin with manikin wipes. Clean the

inside of the oral cavity with manikin wipes. Change the lung
bladders; see: Maintenance - Replacing Lung Bladders.

Note: The manikin airways are not designed for mouth to

mouth rescue breathing or to be disinfected.

CHEST MOVEMENT

Problem

• No chest rise on manikin.

Solution

• Check that manikin is switched ON.

• Check that manikin is not in sleep mode due to inactivity.

Reactivate the manikin.

• Check that awRR is not set to zero on Instructor Application

(Patient Case or scenario)

• Check that airway complications like maximum air resistance or

laryngospasm are not set.

• Check if the internal compressor is switched off. See: Turning the

Internal Compressor Off/On.

• Check that any external compressed air source is switched off

and that the air tube is disconnected from the manikin.

• The internal compressor may have overheated. Wait

approximately 20 minutes for it to cool down. Remove blankets
or covers from the manikin, open the torso skin to facilitate
cooling.

• Chest-rise is set to bilateral (for example if ET-tube is inserted

too far into the bronchia).

• Chest-rise bladder is leaking or tubing to chest-rise bladder is

twisted, kinked or disconnected. Replace chest-rise bladder if it
is leaking, see Maintenance: Replacing Chest Rise Bladders.

• Check air tubing for leakage; check that all connections are intact.

Replacing tubing if necessary.

• Shallow chest movements and the internal compressor runs

continuously. The internal compressor may be worn - contact
Laerdal Technical Service.

BLOOD SYSTEM

Problem

• No blood flow

Solution

• Make sure that the internal reservoir is filled with fluid.

Problem

Too low flow.

Solution

• Clean the blood system

• Check the flow setting in the Instructor Application

• The blood system may need to be re-calibrated. Contact your

local service representative.

• Replace filter

Problem

• Air when bleeding

Solution

• Bleed the blood reservoir empty and refill.

Problem

• Simulated blood is seen leaking from the back of the right leg

during filling or when the manikin is turned off.

Solution

• The internal reservoir may need to be replaced. Contact your

local service representative.

Problem

• Can’t fill the blood reservoir

Solution

• The filter may be clogged. If so, order a replacement fill bottle.

MANIKIN LIMBS

Problems

• Lack of motion in the legs

Possible Solution

• Loosen and re-adjust the hip joint nuts on the inside of the

pelvis. See: “Manikin Setup” for details on how to attach the legs

LUNGS

Problem

• Lungs not functioning properly

Possible Solution

• Check that the Airway resistance is not set to maximum in the

Instructor Application.

• Open the torso and chest plate. Check that the lungs are free to

expand and are not restricted by any cables.

• Check that the lung bladders are properly connected, and that

the tubes are not twisted.

• Check that the lung bladders are in a horizontal position and

inserted correctly. Ensure that the lung compliance O-rings lie
between the folds of the lung bladders.

• Check for flaws or ruptures in the lung bladders.

• Check that the two lung compliance O-rings are fitted correctly.

Replace O-rings if they appear to be damaged.

• Check that there are no obstructions inside the manikin airways

which may block air flow.

• If there is no change when adjusting lung compliance, contact

Laerdal Technical Service.