AMETEK MX Series Rev: AY User Manual

Page 171

Advertising
background image

User Manual - Rev AV

AMETEK Programmable Power

MX Series

171

6.4.1 Output Cal Series II only

300 VAC Range DC Zero:

Program the output to the 300 VAC Range by pressing and
selecting the 300 Range with the shuttle. Program the output to
0.0 volts and 60 Hz. Go to the OUTPUT CAL screen, select the
VOLT ZERO parameter and adjust the output to 0.0 ± 0.005
VDC. Save this value by pressing the ENTER key.

300 VAC Range High Freq DC Zero:

Required for Series II models with –HF option only.

Program the output to the 300 VAC Range by pressing and
selecting the 300 Range with the shuttle. Program the output to
0.0 volts and 820 Hz. Go to the OUTPUT CAL screen, select
the VOLT ZERO parameter and adjust the output to 0.0 ± 0.005
VDC. Save this value by pressing the ENTER key.

To calibrate AC output, set the power source to the AC mode, high voltage range. For MX units
with the –HV optional range, use the 300V range for all calibrations. Select the OUTPUT CAL
entry from the MENU 3 screen. Refer to section 4.2.14 for relevant LCD screen.

300 VAC Range Full-scale:

Before programming the output voltage, first set the ALC mode
to OFF in the UTILITY, VOLT/CURR CONTROL screen. Then
select the 300 VAC Range from the Program 1 screen. Program
the output to 240.0 volts and 60 Hz and close the output relay.
Go to the OUTPUT CAL screen, select the VOLT FS and enter
20000. With the shuttle, adjust this value to obtain 240 ± 0.05
volts at the output as measured with an external DMM. Use the
ENTER key to save this value.

6.4.2 Output Cal 150 V AC Range (Series I only)

To calibrate AC output, set the power source to the AC mode, low voltage range. Select the
OUTPUT CAL entry from the MENU 3 screen. Refer to section 4.2.14 for relevant LCD screen.
For Series II MX units, proceed to section 6.4.3 as there is only one calibration range.

150 VAC Range DC Zero:

[Applies to MX Series I only] Press the PROGRAM key and
select the 150 Range with the shuttle. Program the output to
0.0 volts. Go to the Output Calibration, press the PHASE key to
select the phase to be calibrated and select the VOLT ZERO
parameter. Close the Output Relay by pressing the OUTPUT
ON/OFF key. Monitor the DC output voltage with the external
DVM. With shuttle control adjust the output voltage for 0.0 ±
0.005 VDC. Save this value by pressing the SET key.

150 VAC Range Full-scale:

[Applies to MX Series I only] Press the PROGRAM key and
program 120.0 volts and 60 Hz. Once the output settings are
made, turn on the ALC mode and allow the output on all phases
to settle. Next, proceed to the output calibration screen. Select
the VOLT FS parameter and enter the set voltage in the full-
scale calibration window followed by the ENTER key. A value
between 0.9 and 1.1 will appear in the window. Save this value
by pressing the SET key
.