AMETEK MX Series Rev: AY User Manual

Page 214

Advertising
background image

User Manual - Rev AV

AMETEK Programmable Power

MX Series

214

9.4.6.1 DIPS AND INTERRUPTIONS TEST

Scroll to the DIPS AND INTERRUPTIONS entry using the up and down cursor keys. Press the
ENTER key to select the DIPS AND INTERRUPTIONS menu. The screen will appear as shown
in Figure 9-15.

Figure 9-15: IEC Dips and Interrupts

STATE

This field enables or disables the -411 test mode.

EOS STATE

This field is not available on MX Series systems.

RUN ALL

The RUN ALL selection will cause the following automated test sequence suggested by the
standard to be run:

Step

Output in

% of U

T

No of Cycles

Start angle

(Degrees)

Repeat # times

Delay

between

repeats (s)

1

0

0.5

0

3

10

2

0

0.5

180

3

10

3

0

1

0,45,90

3 at diff ø

10

4

0

5

45,90,135

3 at diff ø

10

5

0

10

90,135,180

3 at diff ø

10

6

0

25

180,225,270

3 at diff ø

10

7

0

50

270,315,0

3 at diff ø

10

8

40

0.5

0

3

10

9

40

0.5

180

3

10

10

40

1

0,45,90

3 at diff ø

10

11

40

5

45,90,135

3 at diff ø

10

12

40

10

90,135,180

3 at diff ø

10

13

40

25

180,225,270

3 at diff ø

10

14

40

50

270,315,0

3 at diff ø

10

15

70

0.5

0

3

10

16

70

0.5

180

3

10

17

70

1

0,45,90

3 at diff ø

10

18

70

5

45,90,135

3 at diff ø

10

19

70

10

90,135,180

3 at diff ø

10

20

70

25

180,225,270

3 at diff ø

10

21

70

50

270,315,0

3 at diff ø

10

Table 9-19: Dips and Interruptions Tests Performed During RUN ALL

The user can change the NOMINAL Ut voltage for this. The RUN ALL Command line will change
to ABORT during the test. Selecting ABORT and pressing the ENTER key will terminate the test
at any time and the output voltage will return to the nominal value.

PHASE SELECTION