DALE601/601E

Operators Manual

1-10

Table 1-2. Components and Controls of the Analyzer (cont.)

Label Name

Function

K

Test Points

0.15

Ω

and 100 µA; provided to check the

Analyzer by clipping onto with the chassis cable
and selecting appropriately, RESISTANCE or
ENCLOSURE LEAKAGE, via the function switch.

L

Chassis
Connector

Provides means for inputting the chassis cable
with its clip for connection to the DUT chassis or
enclosure. With the FUNCTION switch in the
resistance position, the earth resistance is
measured, and on the enclosure position, the
enclosure leakage current is measured.

M

External
Connector

Provided for making external measurement of
resistance, voltage gradient, leakage current or
isolation current of probes and transducers,
when used in conjunction with the chassis cable
for the other reference point.

N

Test Load

Selector switch between IEC601.1 and AAMI

Chassis Cable
(not shown)

A black coil cord with a clamp with black
insulation for measurement of enclosure leakage
and earth resistance. One is supplied with each
Analyzer.

Specifications

The following are specifications for the Analyzer. Please contact your Fluke
Biomedical service representative for more information regarding the device
specifications.

Controls

Function Switch

.............................................

9 position rotary

Lead Switch.........................................................6 position rotary
Neutral Switch .....................................................2 position rocker (Open/Closed)

Polarity Switch.....................................................3 position rocker

(Normal/Off/Reversed)

Leakage Switch ...................................................2 position rocker (Chassis/Earth)