Verifying analyzer performance – Fluke Biomedical DALE601 User Manual

Page 50

Advertising
background image

DALE601/601E

Operators Manual

3-4

Verifying Analyzer Performance

The general performance of the Analyzer can be verified, using the procedures
summarized in Table 3-1 and the parameters in Table 3-2. If the Analyzer does
not perform correctly, contact the Fluke Biomedical Service Center
immediately.

Table 3-1. Performance Verification Parameters

Parameter Procedure

Desired

Reading

Mains Voltage

Use a multimeter.

Volts

Current

Connect a lamp with a
standard incandescent
bulb to the DUT receptacle
and set the OUTLET
switch to NORMAL.

Amps, approximately as
shown in Table 3-2

Resistance

Connect the black chassis
cable with black insulation
to the CHASSIS connector
and clip it to the 0.15

Ω

test point.

0.15 ±0.02

Ω

Enclosure
Leakage

Connect the black chassis
cable with a clamp with
black insulation to the
CHASSIS connector and
clip it to the 100 µA test
point.

100 ±2 µA

Patient Lead
Leakage

Attach a clip lead to the
100 µA test point and the
other end to the LEAD
snap selected.

100 ±2 µA

Patient Aux.
Leakage

Attach a clip lead to the
100 µA test point and the
other end to the LEAD
snap selected.

100 ±2 µA

Advertising
This manual is related to the following products: