Resistance, Xw warning – Fluke Biomedical DALE601 User Manual

Page 45

Advertising
background image

Operation

External

Measurements

2

2-21

2. Set the

FUNCTION

switch to

EXTERNAL

.

MAI

N

S

V

OLTAGE

CURRE

N

T AMPS

EARTH RESISTA

N

CE

EARTH LEAKAGE

E

N

CLOSURE LEAKAGE

PATIE

N

T LEAD LEAKAGE

PATIE

N

T AUX CURRE

N

T

M.A.P.

EXTER

N

AL

fcu11.eps

3. Press the

ISO TEST

switch.

This action applies isolated line voltage between the two sides. The
Analyzer displays the isolation current that flows.

Note

The method for making connection to either side of the isolation
barrier varies with the device to be tested.

XW

Warning

To avoid possible electrical shock or personal injury

take

care when handling the cables. High voltage, 110 percent
of line volts, will be accessible between the two cables
when the LIFT GROUND/ISO TEST switch is in the ISO
TEST position.

Resistance

Resistance measurements between two points are made to verify the integrity
of permanently installed equipment whose ground cannot be broken to
measure the chassis leakage current. These are usually high power devices,
which can have high leakage currents and depend on the bonding of all chassis
to a common point for safety. It is recommended that these devices be powered
off for testing.