Measuring map (patient lead isolation) current – Fluke Biomedical DALE601 User Manual

Page 40

Advertising
background image

DALE601/601E

Operators Manual

2-16

3. Select by the

LEADS

switch to test individual leads.

LEADS

fcu12.eps

4. Make measurements under all combinations of the

OUTLET

switch,

NORMAL

and

REVERSE

; the

NEUTRAL

switch

CLOSED

and

OPEN

;

GROUND CLOSED

and the

LIFT

GROUND

switch

OPEN

;

and the device power turned

ON

and

OFF

.

Under normal conditions, the current is primarily input bias current,
measuring current or leadoff sensing current. The worst case condition is
measured from the individual lead to all other leads connected together.

The single lead carrying the most current is generally considered the
reference lead (RL) because it acts as the return for the other leads.
Therefore, the leads can be taken in groups of four with the common
reference lead (RL) and then summed.

Measuring MAP (Patient Lead Isolation) Current

MAP (Mains on Applied Parts) Current (I

I

) (patient lead isolation current) is

that which would flow into the DUT if the patient were to come into contact
with full line voltage. An example is an electric patient bed that has become
ungrounded and has a short to the frame.

Measurement is made in each individual leads, but a common value is found
for all leads, as well as for the

ALL

position, as this is the measurement for the

isolation of the patient circuit. To assure proper reading, the test should be run
with ground intact. For additional safety, the current is applied to the patient
leads only when the

ISO TEST

switch is pressed.

Advertising
This manual is related to the following products: