Measuring Line (Mains) Voltage

Line Voltage (V

L

) is the mains power supplied by the electrical distribution

system of the hospital. It is shown as a three-wire system of HOT, NEUTRAL
(L2) and GROUND with NEUTRAL, like GROUND, returned to true earth at
entry into the building.

Measurement is made between the HOT and NEUTRAL wires via transformer
coupling to isolate the measuring circuits from the line.

To measure line voltage:

1. Set the

FUNCTION

switch to

MAINS VOLTAGE (

the recommended

starting position).

MAI

N

S

V

OLTAGE

CURRE

N

T AMPS

EARTH RESISTA

N

CE

EARTH LEAKAGE

E

N

CLOSURE LEAKAGE

PATIE

N

T LEAD LEAKAGE

PATIE

N

T AUX CURRE

N

T

M.A.P.

EXTER

N

AL

fcu03.eps

The Analyzer displays the mains voltage with a resolution of 1 V.

2. Set the

OUTLET

switch to the

NORMAL

position and

POWER ON

the

DUT.

3. The Analyzer displays mains voltage, but under the load of the device

being tested. Depending on the DUT operating current and the electrical
supply wiring, the voltage differential may be significant.

Note

The difference between the line voltages with the DUT powered off
and powered on indicates whether the line is adequate for the device.
A large drop may suggest that a dedicated, higher ampacity line
should be run in for the device.