Maintenance and Service

Verifying Analyzer Performance

3

3-5

Table 3-1 Performance Verification Parameters (cont.)

Parameter Procedure

Desired

Reading

M.A.P. (lead
isolation current)

Connect the black chassis
cable with a clamp with
black insulation to the
EXTERNAL connector and
the clip to the lead
selected. Press ISO
TEST.

1000 ±15 %, depending on
line voltage

External; for
leakage current
measurement

Connect the black external
cable with a clamp with
red insulation to the
EXTERNAL connector and
the clip to the 100 µA test
point.

100 ±2 µA

Isolation

Clip the black external
cable with a clamp with
red insulation to the black
chassis cable and press
the ISO TEST switch.

1000 ±15 %, depending on
the line voltage

Calibration

Do not attempt.

Not Applicable

Table 3-2. Current Verification Parameters

Current (Amps)

Bulb (Watts)

120 V Line

230 V line

60 0.50 0.30

75 0.63 0.35

100 0.83 0.45

150 1.25 0.70