Measuring Device Current

2

2-7

4. Check the value under load against the DUT ratings to be sure that the

actual value remains within prescribed limits.

Measuring Device Current

Device Current (I

L

) (instrument current) is that used by the DUT. When

powered on, the DUT should be operated in its various modes to determine the
worst condition to track. Verify that the DUT current is within the current
rating of the Analyzer being used.

Measurement is made in the HOT wire via transformer coupling to ensure that
the total current is measured, because it is possible that the NEUTRAL and
GROUND wires share the return path

To measure device current.

1. Set the

FUNCTION

switch to

CURRENT AMPS

.

MAI

N

S

V

OLTAGE

CURRE

N

T AMPS

EARTH RESISTA

N

CE

EARTH LEAKAGE

E

N

CLOSURE LEAKAGE

PATIE

N

T LEAD LEAKAGE

PATIE

N

T AUX CURRE

N

T

M.A.P.

EXTER

N

AL

fcu04.eps

The Analyzer displays the DUT current to 19.99 A.

2. Set the

OUTLET

switch to

NORMAL

and

L2

to

CLOSED

.

3.

POWER ON

the DUT and place it in its maximum load condition to

obtain the proper reading.

4. Log this data and note changes in values to help detect problems early.