External measurements – Fluke Biomedical DALE601 User Manual

Page 42

Advertising
background image

DALE601/601E

Operators Manual

2-18

LEADS

fcu12.eps

4. Press the

ISO TEST

switch to safely apply the high voltage to the leads.

5. While the

ISO TEST

is energized, read the isolation current in µA.

6. Take measurements for all combinations of the

OUTLET

in

both

NORMAL

and

REVERSE

and

L2

switch

CLOSED

and with the DUT

ON

and

OFF

.

External Measurements

The Analyzer provides the extended capability of making leakage current and
isolation current measurements between selected points, performing as
follows:

An ohmmeter capable of measuring resistance in the m

Ω

range for chassis

bonding measurements.

A low impedance (1000

Ω

) voltmeter as specified by the NFPA, with a

range to 1999 mV.

A 1000 µA ammeter to measure the leakage current between two devices

or between the DUT and a local ground reference. Range is to 1999 µA.

An isolation tester for testing ultrasound probes and other transducers

independently from the device in which it is used.

Note

Additional optional accessory cables must be purchased to perform
these tests.

Advertising
This manual is related to the following products: