Operation

Measuring MAP (Patient Lead Isolation) Current

2

2-17

For this test, the Analyzer provides a specially shielded line voltage secondary
on its power transformer that minimizes internal leakage currents from the
measuring circuit. This avoids the necessity of measuring the current and
subtracting it from all readings. The current is limited with a 120 K

Ω

resistor

for user protection.

XW

Warning

To avoid possible electrical shock or personal injury to the
patient, take care when handling the patient leads. High
voltage, 120/230 volts with respect to earth ground, is
accessible at the patient connections during part of this
test.

To measure lead isolation sink current:

1. Connect the patient leads to the universal lead connectors on the front of

the analyzer.

2. Set the

FUNCTION

switch to

M.A.P

.

MAI

N

S

V

OLTAGE

CURRE

N

T AMPS

EARTH RESISTA

N

CE

EARTH LEAKAGE

E

N

CLOSURE LEAKAGE

PATIE

N

T LEAD LEAKAGE

PATIE

N

T AUX CURRE

N

T

M.A.P.

EXTER

N

AL

fcu10.eps

3. Select the individual lead to be tested by the

LEADS

switch.