Verifying the Power Outlet Connections

Note

This section is not applicable to isolated power systems or to the
601E (230 V version).

Three neon lamps provide indication of the polarity and condition of the outlet
used, as determined by the following chart.

REV

OK

Correct

wiring

Reverse

polarity

Open

ground

Open hot

Hot /

ground

reversed

Note

Lamps do not check open neutral or neutral/ground reversal.

If the line is found faulty, correct the problem before proceeding. If the line
checks OK, plug the device to be tested into the Analyzer's line receptacle.

XW

Warning

To avoid possible electrical shock or personal injury due
to faulty device under test, follow these guidelines:

Do not touch the metal chassis, any accessible
conductive part, or terminal of the DUT with mains
power applied until all applicable tests have been
completed and the product is verified as compliant.

Keep fingers behind guards on supplied test
accessories and/or if necessary remove power from
the DUT before making necessary connections.