Fluke Biomedical DALE601 User Manual

Page 46

Advertising
background image

DALE601/601E

Operators Manual

2-22

Measurement is made with a 10 mA dc current and is highly resistant to ac
currents flowing between the measurement points as a result of chassis leakage
current from either or both devices. However, such leakage current may have a
dc component from rectification within the instrument. This dc current adds
and subtracts from the measurement current, resulting in some error that can be
corrected for by reversing the connections, remeasuring, and averaging the two
readings.

To measure resistance between two points:

1. Connect the two black chassis cables with clamps with black insulation to

the

CHASSIS

and

EXTERNAL

connectors on the top panel of the

Analyzer.

2. Clip the cables onto the two points to be measured.

Note

Do not plug a device into the Analyzer. Ensure that ground
connections do not provide parallel measurement circuits causing
erroneous results.

3. Set the

FUNCTION

switch to

EARTH RESISTANCE

.

MAI

N

S

V

OLTAGE

CURRE

N

T AMPS

EARTH RESISTA

N

CE

EARTH LEAKAGE

E

N

CLOSURE LEAKAGE

PATIE

N

T LEAD LEAKAGE

PATIE

N

T AUX CURRE

N

T

M.A.P.

EXTER

N

AL

fcu05.eps

The display shows the resistance between the two points.