Leakage current – Fluke Biomedical DALE601 User Manual

Page 43

Advertising
background image

Operation

External

Measurements

2

2-19

Leakage Current

Leakage Current and Voltage Gradients between two points are also measured
for permanently installed equipment as additional verification of the integrity
of the installation. These measurements are equivalent, because the
relationship between volts and current across 1000

Ω

is one mV per µA.

To measure leakage current and voltage gradient between two points:

1. Connect the black coil cord cable with the clamp with black insulation to

the

CHASSIS

connector and the black cable with the clamp with red

insulation to the

EXTERNAL

connector on the top panel of the Analyzer.

Note

Do not plug a device into the Analyzer. The measurement circuit must
remain isolated from ground.

Advertising
This manual is related to the following products: