49

• Maximum current: 30 mA for each of the voltages
• Maximum total power: at least 2 W

4.1.5

High Voltage:

Single positive or negative modular HV supply reaching from 50 to 3000 V

Indication of high voltage sign and value by software

Calibration within 1% at maximum of bias voltage

Maximum current 0.5 mA or 0.25 W (e.g. 1000 V and 0.25 mA)

4.1.6 Battery:

Rechargeable Li battery with no memory effect

Deep discharge protected

Battery capacity 2.5 Ah at 20 °C

Measured values of the battery life time

If no detector connected: 16 hours

Expected battery life time with detection systems connected:

Standard IAEA NaI detector (ORTEC/SCIONIX):

14 hours

Bias:

500 V, 100 µA

Preamplifier: 100 mW

Miniature CdTe detection probe (SDP310, RITEC, Latvia): 12 hours
Bias:

400 V, 100 nA

Preamplifier: +12 V/20 mA, -12 V/9.5 mA

Standard planar HPGe detector (e.g. Canberra, Ortec): 6 hours
Bias:

2000V, 1 nA

Preamplifier: +12 V/21 mA, -12 V/7 mA;

+24 V/18 mA, -24 V/8 mA

4.1.7 Computer Interface

RS-232C port (Baud rate: 38400, proprietary software protocol)

4.1.8 Mechanical specification

Size:

155 mm x 95 mm x 45 mm (footprint of palmtop HP 200LX)

Weight:

Box (incl. battery)

ca. 700 g

Line voltage sensing Charger

ca. 320 g

4.1.9 Environmental ratings

Range of storage temperature:

-20 ... +60 degrees C

Range of operational temperatures:

0 - 50 degrees C