Finishing

It is of critical importance to use the correct grinding instruments for finishing and adjusting high-strength glass-ceramics. If

unsuitable grinding instruments are used, chipping of the edges and local overheating may occur (please observe the

Ivoclar Vivadent Flow Chart ”Recommended grinding tools for PS e.max glass-ceramics”.

The following procedure is recommended for finishing pressed-over Straumann Anatomic IPS e.max Abutments:

– The adjustment by grinding of pressed IPS e.max ZirPress restorations is kept to a minimum in the staining technique.

– Separate the sprues with a fine diamond disk under permanent water cooling of the cutting area.

– Overheating of the ceramic must be avoided. Low speed and light pressure is recommended.

– Smooth out the attachment points of the sprue.

– Check the occlusion and articulation and grind in the appropriate adjustments, if necessary.

– Design surface textures.

– To clean the restoration, screw the abutment onto the polishing aid and briefly blast with Al

2

O

3

at 1 bar (15 psi) pressure

and clean with the steam jet. Cover exposed abutment areas with wax!

Use a fine separating disk and continuous water cooling to cut the sprues.

Smooth out the attachment points of the sprue using low speed and light pressure and

design a natural surface structure.

Blast the pressed-on IPS e.max ZirPress with Type 1 Al

2

O

3

at 1 bar (15 psi) pressure with the

polishing aid screwed into place.

Before staining and characterization, thoroughly clean the abutment under running water

and with a steam jet.