Substitute Leakage Test

24

A special limit indicated by a plus sign (+) to the left of the limit, is a
user adjustable limit.

To set your own limit, with this user adjustable limit flashing press
both cursor keys together. The limit as a whole will stop flashing with
the most significant digit flashing. Use the cursor buttons to set a
value for the flashing digit. Press the SELECT button to enable the
next digit in turn to flash. Use the cursor buttons to set a value for this
flashing digit. Then press the SELECT button to move to the next
digit. Repeat these actions until all digits are set to your chosen
value. Press the STORE RESULT button to set the limit.

Press the SELECT button to change to another setting or press the
START/STOP button to perform the test if the other settings are
acceptable.

Substitute Leakage Test

Switch Position

------------

Substitute Leakage

Test Icon Displayed ---

Class 1 (earthed) appliances

Plug the EUT into the tester outlet socket.

Class 2 (unearthed) appliances

Plug the EUT into tester outlet socket, connect the Earth Continuity
lead from its socket to the appliance metal part.

Test Description

The substitute leakage test applies a nominal voltage of 40V AC RMS
to the appliance and is applied between the earth pin and both the
live and neutral pins of the supply plug. For Class II appliances,
connect the Earth Continuity test clip to the appliance as a substitute
for the Earth.