Touch Leakage Test

28

This differential method of determining leakage will show the full
leakage of an appliance in-situ, so if the appliance has an extra earth
point, e.g. a water pipe, then the Tester will show the full and true
appliance leakage current.

Test Duration
The test is continuous and is stopped by either pressing the
START/STOP button or by switching to an un-powered test. In both
cases the test will be stopped and the power to the appliance
removed. Switching to another powered test will stop the previous
test and start the new test without interrupting power to the appliance.

User Selectable Settings

Only the test limit can be set by the user for this test.

Test Limit
With the secondary display flashing and indicating current (mA
displayed), use the cursor keys to step through the test limits
available and stop with the appropriate limit displayed.

A special limit indicated by a plus sign (+) to the left of the limit, is a
user adjustable limit.

To set your own limit, with this user adjustable limit flashing press
both cursor keys together. The limit as a whole will stop flashing with
the most significant digit flashing. Use the cursor buttons to set a
value for the flashing digit. Press the SELECT button to enable the
next digit in turn to flash. Use the cursor buttons to set a value for this
flashing digit. Then press the SELECT button to move to the next
digit. Repeat these actions until all digits are set to your chosen
value. Press the STORE RESULT button to set the limit.

Press the SELECT button to change to another setting or press the
START/STOP button to perform the test if the other settings are
acceptable.

Touch Leakage Test

Switch Position

------------

Touch Leakage

Test Icon Displayed ---