Seaward Europa Pac Plus User Manual

Page 37

Advertising
background image

External Leakage Test

31

Test Duration
The test is continuous and is stopped by either pressing the
START/STOP button or by switching to an un-powered test. In both
cases the test will be stopped and the power to the appliance
removed. Switching to another powered test will stop the previous
test and start the new test without interrupting power to the appliance.

User Selectable Settings

There are no user selectable settings.

External Leakage Test

Switch Position

------------

External Leakage (Clamp)


Test Icon Displayed ---

Warning

Mains voltage applied to appliance

Note

This test requires the use of a current clamp, which is an optional
accessory (see Chapter 5 Accessories).

All Appliances

Plug the current clamp into the appropriate socket. Open the clamp
jaws and clamp around the cable you wish to measure current flow
through. Power is applied to the EUT at the mains socket.

Test Description

The tester will display the current detected by the external clamp. If
this is around a protective conductor, earth leakage will be displayed.
If the clamp is around the phase and neutral conductors then

Advertising