Lead Test

44

Accuracy.........................................................................Indication only

Pass Levels ...................................................................... User defined

Lead Test

Earth Continuity

Test Voltage *. ...............................................4V dc nominal (no load)

Test Current * .....................................+200mA, -200mA (into s/c load)

Range ......................................................................... .00m

Ω - 19.99Ω

Resolution. ................................................................................ ..0.01

Accuracy..................+/- 5% of reading, +/- 2 digits (0.05 - 5.00 ohms)

Pass Levels. ............................................................0.1

Ω, User defined

Insulation

Test Voltage * .............. 500V d.c. or 250V d.c. nominal (0.5M

Ω load)

Short Circuit Current ............................................ 2mA d.c. maximum

Range............................................................................ 00k

Ω - 100 MΩ

Resolution .............................................................0.01 M

Ω (<100 MΩ)

Accuracy...................... .+/- 5% of reading, +/- 2 digits (100k

Ω - 5MΩ)

Pass Levels….. ..….0.25M

Ω, 0.50MΩ, 1.0MΩ, 2.0MΩ, User defined

Power Conductor Continuity

Test .......................................................................................... 40V d.c

Detects............................................................... PASS, FAIL indication

* - Test stimulus based on mains supply of 230V, varies with supply
voltage.