Substitute Leakage Test

43

Substitute Leakage Test

Test Voltage * ................................................................... 40V a.c. o/c

Display Range .......................................................... 0.2 - 20.0 mA a.c.

Resolution. ................................................................................0.01mA

Accuracy...................... +/- 10% of reading, +/- 2 digits (1.00 - 20mA)

Pass Levels. ....... 0.25mA, 0.5mA, 3.5mA, 7mA, 15mA, User defined

Differential Leakage

Range...........................................................................0.0mA - 10.0mA

Resolution.. ...............................................................................0.01mA

Accuracy.......................... +/- 5% of reading, +/- 4 digits (0.2 – 20mA)

Pass Levels ........... 0.25mA, 0.25mA, 0.75mA, 3.5mA, User defined

Touch Leakage

Range.......................................................................... ..0.0mA – 3.5mA

Resolution. ................................................................................0.01mA

Accuracy......................... +/- 5% of reading, +/- 4 digits (0.1 – 3.5mA)

Pass Levels. .......................... 0.25mA, 0.75mA, 3.5mA, User defined

Load Test

Socket .............................................................................230V BS 1363

Measured Voltage.................................... 195 - 300V (Indication only)

Measured Current................................... 0.50 – 16A (Indication only)

Measured Load .............................................0 - 4kVA (Indication only)

External Leakage

Range....................................................................... .0.0mA – 19.99mA

Resolution.. ...............................................................................0.01mA