Seaward Europa Pac Plus User Manual

Page 38

Advertising
background image

IEC Lead Test

32

differential leakage will be displayed. DO NOT connect the clamp
around only the phase or neutral conductors.

Test Duration
The test is continuous and is stopped by either pressing the
START/STOP button or by switching to another test. In both cases
the test will be stopped.

User Selectable Settings

Only the test limit can be set by the user for this test.

Test Limit
With the secondary display flashing and indicating current (mA
displayed), use the cursor keys to step through the test limits
available and stop with the appropriate limit displayed.

A special limit indicated by a plus sign (+) to the left of the limit, is a
user adjustable limit.

To set your own limit, with this user adjustable limit flashing press
both cursor keys together. The limit as a whole will stop flashing with
the most significant digit flashing. Use the cursor buttons to set a
value for the flashing digit. Press the SELECT button to enable the
next digit in turn to flash. Use the cursor buttons to set a value for this
flashing digit. Then press the SELECT button to move to the next
digit. Repeat these actions until all digits are set to your chosen
value. Press the STORE RESULT button to set the limit.

Press the SELECT button to change to another setting or press the
START/STOP button to perform the test if the other settings are
acceptable.

IEC Lead Test

Switch Position

------------

Lead Test


Test Icon Displayed ---