Part 1: Installation

Siting

The 505XY should be situated on a flat surface and is fitted with adjustable

feet to ensure that the unit can be levelled. Adequate area must be given to

allow tubing to pass from the 505L pumphead to the dispensing head of the

robot without kinking or unnecessary bends.

Electrical connection

The 505XY operates on single phase mains electricity only. It is fitted with a

mains voltage selector which must be set to either 120V for 100-120V

50/60Hz supplies or 240V for 220-240V 50/60Hz supplies.
The on/off switch of the 505XY is located at the mains socket as is the fuse.

A spare fuse is incorporated in the fuse holder and should be replaced with

the same value of fuse if used.
A mains cable fitted with a moulded plug is supplied with the 505XY, but if

another plug is to be fitted, the colour coding of the mains lead must be

observed.
The mains cable for 220-240V supplies is coded so that the live lead is

coloured brown, the neutral lead is coloured blue, and the earth lead is

coloured green and yellow.
The mains cable for 100-120V supplies is coded so that the live lead is

coloured black, the neutral lead is coloured white, and the earth lead is

coloured green.
Should the robot fail to operate, check that mains electricity is available at

the unit, that the voltage selector switch is in the correct position and that the

mains fuse is in tact.

Failure to

operate

There are dangerous voltages (at mains potential) inside the unit. If

access is required, isolate the pump from the mains before removing

the cover.

Interfacing with the 505Di

The 505XY has four connector/sockets on the rear panel.
· Mains socket

· 505Di socket
· RS 232 connector
· Floatswitch socket










4