Figure 8-9. Fault 7, High Case Temp

Fault 7, High Case Temp

Figure 8-9 shows the electronic assembly for an
Oxymitter 4000 with a membrane keypad (upper
view) and an Oxymitter 4000 with an LOI (lower
view).

Membrane Keypad

When Fault 7 is detected, The HEATER LED
flashes three times, pauses for three seconds,
and repeats.

1. If the case temperature exceeds [185°F

(85°C)], the temperature control will shut
off and the 4-20 mA signal output will go to
the default value.

2. This signifies that the environment where

the Oxymitter 4000 is installed exceeds the
ambient temperature requirements or that
heat due to convection is causing case
temperature to rise above the limit.

3. Placing a spool piece between the stack

flange and the Oxymitter 4000 flange may
eliminate this problem.

4. If a spool piece does not solve the problem,

relocation is the only solution.

LOI

When Fault 7 is detected, the LOI displays the
"Board Temp Hi" message. Refer to the
comments in paragraphs 1 through 4 above.

37260025

LOI

KEYPAD

Alarms

Board Temp Hi

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

Advertising