6-9

Oxymitter 4000

Temperatures

O2 Temp - Indicates the thermocouple temperature at the sensing cell;
this should always be 1357°F (736°C).

O2 Temp Max - Maximum temperature the cell has seen. (Some process
temperatures can exceed the 1357°F (736°C) setpoint temperature, and
this will indicate this condition.)

Board Temp - The temperature inside the Oxymitter electronics housing
185°F (85°C), is the max.

Board Temp Max - This is the maximum temperature that the electronics
has experienced over time.

Voltages

The raw mV signals feeding the temperature indications listed in the
previous paragraph.

Output Values

Indication of the current readings for O

2

and mA.

LOI INSTALLATION

The LOI module connects to the top of the electronic assembly in the
electronics housing. There are four matching connectors (Figure 6-5) on the
back of the LOI module that allow the user to orient (rotate) the LOI as
desired.

Figure 6-5. LOI Module

Connectors

37260055

Connector

Receptacles

LOI Module

Rear View