Oxymitter 4000, Alarms o2 temp low, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 106

Advertising
background image

8-14

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Oxymitter 4000

Figure 8-10. Fault 8, Low Heater Temp

Fault 8, Low Heater Temp

Figure 8-10 shows the electronic assembly for
an Oxymitter 4000 with a membrane keypad
(upper view) and an Oxymitter 4000 with an LOI
(lower view).

Membrane Keypad

When Fault 8 is detected, the HEATER LED
flashes four times, pauses for three seconds,
and repeats.

1. The low heater temperature alarm is active

when the thermocouple reading has
dropped below 28.6 mV.

2. If the thermocouple reading continues to

ramp downward for one minute and does
not return to the temperature set point of
approximately 29.3 mV, then an Open
Heater fault will be displayed.

3. Power down the electronics. Remove the

electronic assembly per "Electronic
Assembly Replacement" in Section
9:
Maintenance and Service. Usin
g a
multimeter, measure the resistance across
the terminals of heater connector, J8.

4. If the heater is good, the reading will be

approximately 70 ohms. If the heater is
open, see "Heater Strut Replacement" in
Section 9: Maintenance and Service.

LOI

When Fault 8 is detected, the LOI displays the
"O2 Temp Low" message. Refer to the
comments and procedures in paragraphs 1
through 4 above.

37260026

LOI

KEYPAD

Alarms

O2 Temp Low

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

Advertising