Figure 9-8, Oxymitter 4000, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 131

Advertising
background image

Instruction Manual

IM-106-340, Rev. 4.0
May 2006

9-15

Oxymitter 4000

Figure 9-8. Heater Strut

Assembly

13. For units with remote electronics, install the probe head as follows:

a. See Figure 9-9. Make sure that the O-ring is in good condition. Seat

the O-ring in the mating groove of the probe.

b. Insert the probe signal cables into the housing.
c. Turn the conduit ports of the housing to the CAL and REF gas ports

side of the probe and position the housing on the probe.

d. Install and tighten four screws (2, Figure 9-4).
e. Reconnect the probe signal cables to the probe signal and heater

wire connectors, Figure 9-9. Make sure the connectors are secure.

f. Install and tighten cover.

14. Follow the instructions in "Removal and Replacement of Probe" to

install the Oxymitter 4000 into the stack or duct.

Ceramic Support Rod

Cell Flange

Ceramic

Diffuser

Assembly

V-Deflector

Heater

Wire

Loop

22220050