Oxymitter 4000, Alarms o2 cell bad, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 111

Advertising
background image

Instruction Manual

IM-106-340, Rev. 4.0
May 2006

8-19

Oxymitter 4000

Figure 8-15. Fault 13, Invalid Slope

Fault 13, Invalid Slope

Figure 8-15 shows the electronic assembly for
an Oxymitter 4000 with a membrane keypad
(upper view) and an Oxymitter 4000 with an LOI
(lower view).

Membrane Keypad

When Fault 13 is detected, the CALIBRATION
LED flashes once, pauses for three seconds,
and repeats.

1. During a calibration, the electronics

calculates a slope value. If the value of the
slope is less than 35 mV/dec or more than
52 mV/dec, the slope alarm will be active
until the end of the purge cycle.

2. See "Calibration with Keypad" in Section 9:

Maintenance and Service. Verify the
calibration by carefully repeating it. Ensure
the calibration gases match the calibration
gas parameters. If you attach a multimeter
to TP1+ and TP2-, sample gas
measurements are:

8% O

2

.

23 mV

0.4% O

2

.

85 mV

3. Power down the Oxymitter 4000 and

remove it from the stack.

4. Replace the cell per "Cell Replacement" in

Section 9: Maintenance and Service.

LOI

When Fault 13 is detected, the LOI displays the
"O2 Cell Bad" message. Refer to the comments
and procedures in paragraphs 1 through 4
above.

37260031

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

LOI

KEYPAD

Alarms

O2 Cell Bad

Advertising