Section 8 troubleshooting, Overview, Overview -1 – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 93: Fer to section 8: troubleshooting, S (section 8: troubleshooting, To section 8: troubleshooting, to, Oxymitter 4000

Advertising
background image

Instruction Manual

IM-106-340, Rev. 4.0
May 2006

Oxymitter 4000

http://www.processanalytic.com

Section 8

Troubleshooting

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8-1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8-3
Alarm Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8-3
Alarm Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8-4
Identifying and Correcting Alarm Indications . . . . . . . . . page 8-5
Calibration Passes, but Still Reads Incorrectly . . . . . . . . page 8-22

OVERVIEW

While the Oxymitter 4000 electronics provides a significant number of
diagnostic alarms to assist in troubleshooting potential problems, it is good to
place these alarms in perspective with respect to the instrument's operating
principles:

When the Zirconium Oxide sensing cell is heated to its setpoint [1357°F
(736°C)], the cell will generate a voltage that represents the difference
between the process O

2

% and the reference O

2

% inside the probe (20.95%

O

2

ambient air).

Test points, Figure 8-1, are provided to read the raw millivolt value generated
by the thermocouple that controls the cell temperature and also the raw cell
signal.

The cell temperature at test points 3 and 4 should always be stable at
approximately 29 to 30 millivolts, which represents the [1357°F (736°C)]
setpoint temperature.

When flowing calibration gasses, the raw cell millivolt value at test points 1
and 2 should represent the levels on the chart in Figure 8-1. Note that the raw
cell millivolt value increases logarithmically as the O

2

concentration

decreases.