Oxymitter 4000, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 35

Advertising
background image

Instruction Manual

IM-106-340, Rev. 4.0
May 2006

1-17

Oxymitter 4000

Signals

Analog Output/HART

4-20 mA isolated from power supply, 950 ohms
maximum load

Logic I/O

Two-terminal logic contact configurable as either an
alarm output or as a bi-directional calibration
handshake signal to IMPS 4000 or SPS 4001B,
self-powered (+5 V) in series with 340 ohms
Conduit ports — 3/4 in.-14 NPT (for analog output and
logic I/O signal lines)

Power Requirements:

Probe Heater

175 W nominal

Electronics

10 W nominal

Maximum

500 W

Oxymitter Specifications