When a calibration has been initiated, the value at TP5 and TP6 is the
% O

2

seen by the cell.

Oxygen levels, as seen on the multimeter, are:

8.0% O

2

= 8.0 VDC

0.4% O

2

= 0.4 VDC

2. HART/AMS.
3. Model 751. The loop-driven LCD display.

LOGIC I/O

This two-terminal logic contact can be configured either as a solid-state
relay-activated alarm or as a bi-directional calibration handshake signal to an
IMPS 4000 or SPS 4001B. The configuration of this signal depends on the
setting of the LOGIC I/O PIN MODE via HART/AMS or LOI. The ten different
modes available are explained in Table 3-1.

Table 3-1. Logic I/O

Configuration (as set at

HART/AMS or LOI)

*The default condition for an Oxymitter 4000 without an IMPS 4000 or SPS 4001B.

**The default condition for an Oxymitter 4000 with an IMPS 4000 or SPS 4001B.

Alarm

When configured as an alarm, this signal alerts you to an out-of-spec
condition. The output is 5 V in series with a 340 ohm resistor. For optimum
performance, Emerson Process Management recommends connecting the
output to a Potter & Brumfield 3.2 mA DC relay (P/N R10S-E1Y1-J1.0K).

Of the ten modes in Table 3-1, mode 1 through mode 7 are the alarm modes.
The factory default is mode 5 for Oxymitter 4000 units without an IMPS 4000
or SPS 4001B. In this mode, the output will signal when a unit alarm or a
CALIBRATION RECOMMENDED indication occurs.

Mode

Configuration

0

The unit is not configured for any alarm condition.

1

The unit is configured for a Unit Alarm.

2

The unit is configured for Low O

2

.

3

The unit is configured for both a Unit Alarm and Low O

2

.

4

The unit is configured for a High AC Impedance/CALIBRATION
RECOMMENDED.

5*

The unit is configured for both a Unit Alarm and a High AC
Impedance/CALIBRATION RECOMMENDED.

6

The unit is configured for both a Low O

2

and High AC Impedance/CALIBRATION

RECOMMENDED.

7

The unit is configured for a Unit Alarm, a Low O

2

, and a High AC

Impedance/CALIBRATION RECOMMENDED.

8**

The unit is configured for a calibration handshake with IMPS 4000 or SPS 4001B.
CALIBRATION RECOMMENDED will initiate the calibration cycle.

9

The unit is configured for a calibration handshake. CALIBRATION
RECOMMENDED will not initiate the calibration cycle with the IMPS 4000 or
SPS 4001B.