Oxymitter 4000 test points, Remote powered loop lcd display (optional), Oxymitter 4000 – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 82: Instruction manual

Advertising
background image

Oxymitter 4000

6-10

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

OXYMITTER 4000 TEST
POINTS

Refer to Figure 6-6. System test points are located on the board below the
LOI module. Test points 1 through 6 allow you to monitor with a multimeter:
the heater thermocouple, the O

2

cell millivolt, and the process O

2

.

• TP1 and TP2 monitor the oxygen cell millivolt output which equates to

the percentage of oxygen present.

• TP3 and TP4 monitor the heater thermocouple.
• TP5 and TP6 monitor the process gas or the calibration gas parameter.

REMOTE POWERED
LOOP LCD DISPLAY
(OPTIONAL)

Refer to Remote Powered Loop LCD manual for calibration and operation.


Figure 6-6. Oxymitter 4000 -

Test Points

37260037

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6