Oxymitter 4000, Alarms calib failed, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 113

Advertising
background image

Instruction Manual

IM-106-340, Rev. 4.0
May 2006

8-21

Oxymitter 4000

Figure 8-17. Fault 15, Last Calibration Failed

Fault 15, Last Calibration Failed

Figure 8-17 shows the electronic assembly for
an Oxymitter 4000 with a membrane keypad
(upper view) and an Oxymitter 4000 with an LOI
(lower view).

Membrane Keypad

When Fault 15 is detected, the CALIBRATION
LED flashes three times, pauses for three
seconds, and repeats.

1. The last calibration failed alarm activates

when the slope and constant values
calculated are out of range and the unit
reverts to using the previous calibration
values.

2. The cell should be replaced. See "Cell

Replacement" in Section 9: Maintenance
and Service, for cell re
placement
instructions.

LOI

When Fault 15 is detected, the LOI displays the
"Calib Failed" message. Refer to the comments
in paragraphs 1 and 2 above.

37260033

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

LOI

KEYPAD

Alarms

Calib Failed