Figure 3. Oxymitter 4000 without

SPS 4001B Wiring Diagram

QUICK REFERENCE
GUIDE MANUAL
CALIBRATION
INSTRUCTIONS

Performing a Manual Calibration with a Membrane Keypad

1. Place the control loop in manual.
2. Press the CAL key. The CAL LED will light solid.
3. Apply the first calibration gas.
4. Press the CAL key. When the unit has taken the readings using the first

calibration gas, the CAL LED will flash continuously.

5. Remove the first calibration gas and apply the second calibration gas.
6. Push the CAL key. The CAL LED will light solid. When the unit has

taken the readings using the second calibration gas, the CAL LED will
flash a two-pattern flash or a three-pattern flash. A two-pattern flash
equals a valid calibration, three-pattern flash equals an invalid
calibration.

7. Remove the second calibration gas and cap off the calibration gas port.
8. Press the CAL key. The CAL LED will be lit solid as the unit purges.

When the purge is complete, the CAL LED will turn off.

9. If the calibration was valid, the DIAGNOSTIC ALARMS LEDs indicate

normal operation. If the new calibration values are not within the
parameters, the DIAGNOSTIC ALARMS LEDs will indicate an alarm.

10. Place the control loop in automatic.

AC L1

AC N

+

+

-

-

4-20

Terminal

Block

AC Line

Voltage Port

Signal

Port

Left Side of

Oxymitter 4000

4-20 mA

Signal

29770003

AC Terminal

Cover

Line Voltage

(85 to 264 VAC)

Ground
Lugs

Logic I/O