10

Setup

Measure screen

The Measure screen contains fields for specifying

the method of data point collection, specifying the

minimum number of points required to measure dif-

ferent geometries, locking the reference frame, se-

lecting distance presentations and configuring point

filtration.

Annotation

The annotation type defines the method of data

point collection. The annotation choice field can be

toggled between Backward and Forward annotation types.

Backward annotation imposes no limit on the number of data points that can be collected by the user when

measuring a feature. For example, when backward annotation is used, a circle can be measured with as

few as the minimum number of points shown in the Min Circle data field, or as many as the user wishes to

collect up to a maximum of 99 points. When backward annotation is used, the user must press the Finish

key to complete the measurement and store measurement data in the feature list.

Forward annotation limits the number of data points collected in a feature measurement to the minimum

shown in the corresponding Min data field. For example, when forward annotation is used and the Min

Circle value is 3, the circle measurement will automatically be completed when 3 points have been col-

lected. No user interaction will be required to store the measurement data in the feature list.

Touch the Annotation choice field to toggle the choice be-

tween Backward and Forward annotation.

Minimum points required for a feature measurement

Each feature type is given a minimum data point requirement. The default minimums are the geometric

minimums necessary to define the corresponding feature type. For example, a minimum of 3 points is

required to define a circle, 2 for a line and 1 for a point. The minimum number required by the system

can be increased (up to 99) to improve the accuracy of measurements. This is most useful when Forward

annotation is used and the measurement data are automatically

limited to the specified minimum.

To define the minimum data points required by the system to

complete a feature measurement, touch the corresponding Min

data field and enter the desired minimum number using the

number keys.

The minimum points for a circle are

changed from 3 to 8

Probe Screen and Measure Screen

Advertising